Authors | مهدی شبانی نوش آبادی,فهیمه طاهرنژاد جوزمی |
---|---|
Journal | RSC ADV |
Page number | 56255 |
Volume number | 5 |
IF | 3.84 |
Paper Type | Full Paper |
Published At | 2015-06-11 |
Journal Grade | Scientific - research |
Journal Type | Electronic |
Journal Country | Iran, Islamic Republic Of |
Journal Index | ISI ,SCOPUS |
Abstract
Cyclic voltammetry (CV), square wave voltammetry (SWV), electrochemical impedance spectroscopy (EIS) and double potential step chronoamperometry were used to investigate the electrochemical oxidation of glutathione (GSH) at a chemically modified electrode prepared by incorporating acetylferrocene (AF) and NiO nanoparticles (NiO/NPs) into a carbon paste matrix. For the mixture containing GSH and vitamin B6, the peak potentials were well separated from each other. Under optimal conditions in cyclic voltammetry, linear ranges spanned a GSH concentration from 0.2mM to 350.0mM and the detection limit was 0.09 mM at a signal-to-noise ratio of 3. The novel sensor was successfully used for the voltammetric determination of the GSH in real samples with satisfactory results.
tags: NiO nanoparticle, glutathione , vitamin B6