Detection and dosimetry of gamma ray emitted from thallium-201 and technetium-99m based on chemiluminescence technique

AuthorsMostafa Shourian , Hassan Tavakoli , Hedayatollah Ghourchian , Hossain Ali Rafiee Pour
JournalRadiation Measurement
Page number906-910
Volume number45
Paper TypeFull Paper
Published At2010
Journal GradeISI
Journal TypeTypographic
Journal CountryUnited Kingdom

Abstract

This report describes the detection and dosimetry of gamma ray emitted from Thallium-201 (201Tl) and
Technetium-99m (99mTc) based on chemiluminescence technique. H2O2 produced by two gamma emitter
radioisotopes of 201Tl and 99mTc were quantitatively measured by chemiluminescence method. Upon
producing H2O2 in a luminol alkaline solution, in the presence of diperiodatocuprate, as catalyst
a chemical reaction was accrued and consequently the emitted light was measured. The determined
H2O2 concentration was correlated with the gamma ray detection and dosimetry. The sensitivity of
chemiluminescence technique for 201Tl and 99mTc dosimetry was determined to be 0.20 and 0.08 MBq/l
(Mega Becquerel per liter) respectively (R.S.D. ¼ %5, N¼ 3). The plotted calibration curves showed
detection limits of 3.24 and 1.76 MBq/l for 201Tl and 99mTc, respectively.