نویسندگان | محمد مظلوم اردکانی,پونه ابراهیمی,محمد رضا منصور نیا |
---|---|
نشریه | Russian Journal of Electrochemistry |
شماره صفحات | 1066 |
شماره مجلد | 44 |
ضریب تاثیر (IF) | ثبت نشده |
نوع مقاله | Full Paper |
تاریخ انتشار | 2008-09-30 |
رتبه نشریه | علمی - پژوهشی |
نوع نشریه | الکترونیکی |
کشور محل چاپ | ایران |
نمایه نشریه | SCOPUS ,JCR |
چکیده مقاله
A new salicylate-selective electrode based on the complex (2,3;6,7;10,11;14,15-tetraphenyl- 4,9,13,16-tetraoxo-1,5,8,12-tetraazacyclohexadecane) copper(II) [CuL] as the membrane carrier was developed. The electrode exhibits a good Nernstian slope of –60.9 ± 1.0 mV/decade and a linear range from 1.0 × 10 –6 to 1.0 × 10 –1 M for salicylate. The detection limit is 5.0 × 10 –7 M. The electrode has a fast response time (5–15 s) and can be used for more than three months. The selective coefficients were determined by the fixed interference method (FIM) and separate solution method (SSM). The salicylate-selective electrode could be used in the pH range 3.5–10.5. It was employed as an indicator electrode for direct determination of salicylate in pharmaceutical samples.
tags: salicylate, salicylate-selective electrode, potentiometry, ionophore.